HardyCHROM CRE, selective and differential culture medium for antimicrobial susceptibility testing, Product Catalog No. G323

FDA Recall Z-2041-2025 · Posted 2025-06-27

Recalling FirmHardy Diagnostics
Recall StatusOpen, Classified
Product CodeJSO
Date Initiated2025-05-06
Product Quantity4300 ea

Reason for Recall

Identified lots failed incoming quality control testing. Lots exhibited breakthrough growth of Klebsiella pneumoniae.

Firm Action

On May 6, 2025, the firm began notifying customers via "URGENT Medical Device Recall Notification" letters. Customers were instructed to discard any remaining stock and to destroy it according to local regulations. If product was further distributed, notify the consignees who received the product. Please retain this letter or post the letter where the product is stored to ensure the users are aware of this defect. For replacements or credit, please contact the firm's Technical Services Department at (800) 266-2222, option 2 or via email at [email protected]. Replacements will be sent out at no-charge to your laboratory or a credit will be applied to your invoice upon request. If you have any questions, call 805-346-2766 or email: [email protected].

Root Cause

Nonconforming Material/Component

Distribution Pattern

US domestic distribution to the following states: AZ, CA, CO, FL, GA, IL, MD, ME, MI, MN, MO, NC, NE, NV, NY, PA, PR, TX, UT, VA, WI and WV.