Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

B-Hydroxybutyrate LiquiColor, Item code CH2440058 - Product Usage: Product is used for the quantitative determination of B-Hydroxybutyrate in human serum or plasma.

FDA Recall Z-2049-2020 · Posted None

Recalling FirmCardinal Health Inc.
Recall StatusTerminated
Product CodeJIN
Date Initiated2020-04-20
Date Terminated2020-07-23

Reason for Recall

Three products requiring refrigerated storage conditions were incorrectly stored outside of approved temperature ranges. Using these temperature abused products may result in misdiagnosis of the patient.

Firm Action

On April 20, 2020, the firm distributed Urgent Medical Device Recall letters to affected customers. Customers were informed that the product had been incorrectly stored and that there was a potential for misdiagnosis. Action Required: 1. CHECK all storage and usage locations to confirm whether you have any units of the affected product codes and lot numbers containing the labeling outlined in the letter. 2. SEGREGATE and QUARANTINE all on-hand affected product. 3. PLEASE RETURN the letter's acknowledgment form via facsimile (847-689-9101 or 614-652-9648) or email ([email protected]) and indicate the product code, lot and quantity of product youve quarantined or discarded. Please respond regardless of whether or not you have affected product. 4. NOTIFY any customers to whom you may have distributed, or forwarded product affected by this recall. Your notification to your customers may be enhanced by including a copy of this recall notification letter. 5. CONTACT the appropriate Customer Service group to arrange for return and credit/replacement of any affected product Monday Friday between 8:00am - 5:00pm EST: " Hospital800-964-5227 " Federal Government800-444-1166 " Distributor800-635-6021 " All other Customers888-444-5440

Root Cause

Storage

Distribution Pattern

US Nationwide distribution NC, KS, and AL.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑