ACUSON S2000 ultrasound system Manufactured by Siemens Medical Solutions USA, Inc. Business Unit Ultrasound, 1230 Shorebird Way Intended for the following applications: Fetal, abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/GYN, Cardiac, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal and Peripheral Vascular Applications.
FDA Recall Z-2051-2011 · Posted 2011-04-22
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc.
Recall StatusTerminated
Product CodeIYN
Date Initiated2011-03-15
Date Terminated2012-06-18
Product Quantity1533 units
Reason for Recall
The Firm discovered a software error on the Accuson S2000 ultrasound systems affecting the measurement results in the patient report.
Firm Action
Siemens issued a Customer Safety Advisory Notification letter to all customers on March 15, 2011, regarding a software error on the ACUSON S2000 ultrasound systems affected measurement . They are in the process of releasing a software update that will correct this issue. Siemens recommends reviewing the patient report at the end of each study to confirm that the report contains only measurements relevant to the exam just performed.
No patient injury has been reported.
Further questions regarding this recall please call (650) 694-5398.
Root Cause
Software design
Distribution Pattern
Worldwide Distribution including USA, states of AZ, VA, CA, NE, IL, MN, TN, FL, NM, and DC
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.