Prodisc C SK U.S. IMPLANT EXTRA LARGE 6MM. Model Number: PDSXL6. Total cervical disc replacement.
FDA Recall Z-2051-2026 · Posted 2026-05-01
Advertisement
Recalling FirmCentinel Spine, Inc.
Recall StatusOpen, Classified
Product CodeMJO
Date Initiated2026-03-20
Product Quantity30 units
Reason for Recall
Product labeling mix up. The prodisc C SK U.S. Implant Extra Large 6mm product was labeled as a 5mm and prodisc C SK U.S. Implant Extra Large 5mm product was labeled as a 6mm.
Firm Action
Centinel Spine notified distributors on about 03/20/2026 via telephone calls and emailed letter. Surgeons were sent letters via FedEx on 03/24/2026 and by email on 03/25/2026 & 03/26/2026. Distributors were instructed to identify any affected units on hand, provide photographs documenting such, and arrange for the return of all affected units. Additionally, distributors have been instructed to diligently check prodisc C SK inventory prior to any scheduled surgery to ensure that the recalled lot numbers are not available for implantation. This includes both consignment and loaner inventory. Acknowledgment of the letter has been requested with return envelopes provided.
For surgeons, they were instructed to monitor any patient(s) that have received the recalled device for any unexpected device-related symptoms throughout the regularly scheduled check-in procedures.
Centinel Spine intends to remove from the market all remaining implants within the lots effected by this recall and issue a replacement product with corrected labeling for any remaining products that are subject to this recall.
Root Cause
Labeling mix-ups
Distribution Pattern
US Nationwide distribution in the states of AZ, CA, GA, LA, MO, NY, TN, TX.