Linear Straight Broach Handle (Rasp Handle) (Greatbatch Part # 510397; DJO Surgical Inc., # 803-03-159) and EZ Clean" Linear Broach Handle Greatbatch Part # T11824, DJO Surgical Inc. # 803-03-059).

FDA Recall Z-2055-2017 · Posted None

Recalling FirmGreatbatch Medical
Recall StatusTerminated
Product CodeHTQ
Date Initiated2017-04-23
Date Terminated2019-03-07
Product Quantity205

Reason for Recall

Greatbatch Medical has voluntarily initiated a Recall for the Linear Straight Broach Handle (Rasp Handle) and EZ Clean" Linear Broach Handle. Greatbatch Medical has received reports that that the product is breaking at the welds and that the locating pin at the distal end of the instrument is breaking off during use.

Firm Action

Greatbatch sent an Urgent Medical Device Recall letter dated April 23, 2017. The letter described the product being recalled and described Reason For Recall, Risk To Health, Replacement Devices, & Instructions to DJO Surgical Inc. Advised consignee to quarantine and return the affected product. If the product was further distributed they should contact the end users and share the recall information with them. Requested consignees to complete and return the "Field Recall Response Form" to Stericycle at [email protected] or fax it to 877-479-8063. For questions contact your Greatbatch Medical representative, or contact us at 1-763-951-8235, or e-mail us at [email protected]. For further questions, please call (763) 951-8184.

Root Cause

Nonconforming Material/Component

Distribution Pattern

US Distribution to the state of : TX