R3 TM CERAMIC ACETABULAR LINER, BIOLOX FORTE, 36 MM O.D. 54 MM O.D., REF 71338954, STERILE R, QTY (1), Smith & Nephew, Inc., Memphis, TN 38116, 2017/06
FDA Recall Z-2056-2011 · Posted 2011-04-22
Advertisement
Recalling FirmSmith & Nephew Inc
Recall StatusTerminated
Product CodeMRA
Date Initiated2011-03-11
Date Terminated2012-02-08
Product Quantity571 units
Reason for Recall
During the manufacturing process for several batches of R3 Ceramic Liners, the titanium rings were pressed onto the ceramic component with a higher force than allowed by manufacturing specifications. This has the potential to result in lower than expected strength for the liners.
Firm Action
All affected Smith & Nephew Distributors were notified via email and the hospitals were notified by letter delivered by Fed-Ex on 3/11/2011. They were instructed to immediately cease distribution, quarantine affected product for return, if product further distributed, identify customers and immediately notify them of recall.
Please contact Mr. David Archer, Group Director, at 701.396.2121, ext. 5373, if you have questions.
Root Cause
Employee error
Distribution Pattern
AK, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, MA, MD, ME, MI, MN, MO, NC, NH, NJ, NV, NY, OH, OR, PA, SC, TN, TX, WA, and WI
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.