Terminated

OrthoPediatrics PediLoc Locking Plate System, 4.5mm Contour Locking Compression Femur 8 Hole Plate Right. Used for pediatric patients as indicated for pelvic, small and long bone fractures.

FDA Recall Z-2057-2012 · Posted 2012-07-20

Recalling FirmOrthoPediatrics Corp
Recall StatusTerminated
Product CodeHRS
Date Initiated2012-06-14
Date Terminated2013-04-30
Product Quantity15 devices

Reason for Recall

The device is a Left 8 hole plate but is incorrectly labeled and etched as a Right device.

Firm Action

On 15-June-2012, a follow-up, URGENT Recall of MIS-BRANDED PediLoc Locking Plate e-mail was sent to all Distributors and direct representatives to provide them with a written notification of the voluntary recall. The e-mail identified the affected product and the reason for the recall. It also instructed them to quarantine all identified devices from the affected lot number immediately. A certified Voluntary Recall notice was also sent as a follow up along with a return prepaid pouch for the return of the recalled device(s) and Mandatory Reply Form.

Root Cause

Error in labeling

Distribution Pattern

Worldwide Distribution -- USA, including states of IN, GA, MI, FL, MO, OH, and SC and country of Australia.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →