Recall StatusOpen, Classified
Date Initiated2025-05-28
Product Quantity5,230 units
Reason for Recall
Ultrasound transducer devices were refurbished beyond their useful life.
Firm Action
An URGENT Medical Device Recall notification dated 5/28/25 was mailed to consignees. The notification informs consignees that Philips Ultrasound will be replacing transducers that were refurbished beyond their useful life with a new transducer. Consignees may continue to use affected devices while awaiting replacement; consignees should use devices in accordance with their intended use and per the instructions provided int he User Manual. Users should perform visual inspection and electrical safety testing as indicated in the User Manual; if a device does not pass inspection, it should be removed from clinical use. Consignees are asked to return the completed Response Form and circulate the provided notification to all users to ensure awareness. A copy of the recall notification should be kept with affected transducers until their replacement. Consignees with any questions can contact the Customer Care Solutions Center at 1-800-722-9377.
Additional consignees were identified and notified via URGENT Medical Device Recall Notification on 9/5/25.
Root Cause
Process control
Distribution Pattern
Domestic: Nationwide Distribution; Foreign: Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Morocco, Netherlands, New Caledonia, New Zealand, Nicaragua, Oman, Philippines, Poland, Portugal, Romania, Russian Fed., Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, T¿rkiye, United Kingdom, United Arab Emirates, Venezuela, Vietnam.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.