Synthes Trauma Nail System. The devices are indication for bone fixation.

FDA Recall Z-2058-2013 · Posted 2013-08-26

Recalling FirmSynthes USA HQ, Inc.
Recall StatusTerminated
Product CodeHSB
Date Initiated2012-11-02
Date Terminated2016-09-27
Product Quantity1,011,783

Reason for Recall

Recall was initiated due to the possibility that the outer pouch was compromised. The product is packaged in two pouches; a sterile, internal pouch within an external pouch. It is the outer, external pouch that may be compromised.

Firm Action

The firm initiated their recall of this product on November 2, 2012 by sending an "URGENT NOTICE: MEDICAL DEVICE RECALL" letter to all consignees. The letter explained the problem and gave instructions for consignees to examine their inventory and urged consignees to return the identified devices. The letter stated If returning the product would adversely impact their ability to provide necessary medical care to patients, they should re-sterilize the product per Pre-Vacuum Sterilization instructions contained in the Instruction for Use.

Root Cause

Packaging change control

Distribution Pattern

Nationwide distribution.