Bone Dowell Harvest Tubes
FDA Recall Z-2059-2017 · Posted 2017-04-25
Advertisement
Recalling FirmZimmer Biomet, Inc.
Recall StatusTerminated
Product CodeLXH
Date Initiated2017-04-12
Date Terminated2017-11-03
Product Quantity49
Reason for Recall
mislabeled as 8mm tube but it is
actually 9 mm in size. A 9mm subcomponent was erroneously substituted and etched as 8mm for the actual
8mm subcomponent part.
Firm Action
On 4/12/2017 URGENT MEDICAL DEVICE RECALL REMOVAL notifications were sent to the affected consignees via e-mail. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification.
Root Cause
Nonconforming Material/Component
Distribution Pattern
No US distribution. JAPAN NETHERLANDS SINGAPORE SWITZERLAND
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.