Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Savaria SL-1000 Stairlift

FDA Recall Z-2060-2017 · Posted None

Recall StatusTerminated
Product CodePCD
Date Initiated2017-04-12
Date Terminated2018-03-05
Product Quantity1425 in the United States 734 in Canada

Reason for Recall

The electrical controller, combined with motor, causes feedback, damaging a component on the control board, starting a fire.

Firm Action

The firm sent a letter to dealers dated July 26, 2016. The firm will contact consignees (distributors) via e-mail and phone.

Root Cause

Component design/selection

Distribution Pattern

Distribution in the United States and Canada

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑