MAC VU360, Model Number 2030360-001, Electrocardiograph

FDA Recall Z-2060-2019 · Posted None

Recalling FirmGE Healthcare, LLC
Recall StatusTerminated
Product CodeDQK
Date Initiated2019-06-28
Date Terminated2021-04-13
Product Quantity369 units

Reason for Recall

Two issues: 1) The MAC VU360 system may intermittently display an incorrect patient ID or visit number on the screen after scanning the patient s barcode. This may result in the ECG report being assigned to the incorrect patient. 2) The MAC VU360 may have the incorrect patient demographics appear on the patient banner. This may result in the ECG report being assigned to the incorrect patient.

Firm Action

The firm issued an Urgent Medical Device Correction letter on 06/28/2019 to tis consignees by mail. This letter informed the customer of the issues and provides instructions for continued use of the system. GE Healthcare will correct the system by upgrading the software.

Root Cause

Software design

Distribution Pattern

Worldwide