Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Apex ARC Hip Stem, HA Coated, Size 5; Product Code: HC-21005-HA Product Usage: The Apex ARC Hip Stem is the femoral component of a primary or revision total hip replacement. This femoral hip stem is intended for press-fit application and is provided sterile for single use. They are manufactured from titanium alloy. The stem has a proximal porous coating of titanium plasma spray under a coating of hydroxyapatite (HA)

FDA Recall Z-2064-2015 · Posted 2015-07-14

Recalling FirmOmnilife Science Inc.
Recall StatusTerminated
Product CodeLZO
Date Initiated2015-06-03
Date Terminated2015-09-08
Product Quantity146 devices

Reason for Recall

Product may breach the inner and outer sterile pouches during shipping or handling.

Firm Action

All product is consigned to US Agents until implanted or returned to OMNI. All US Agents holding the potentially affected inventory were notified that the product is being recalled via letter on June 3, 2015. The letter instructs them to immediately discontinue use of the product and contact OMNI Customer Service at 800-448-6664 to request a return materials authorization, and to return the Acknowledgement and Receipt form by FAX to 508-819-3390 or email to [email protected].

Root Cause

Packaging

Distribution Pattern

US Nationwide Distribution in the states of: UT, CA, FL, TX, MA, GA, OK, VA, CT, OK, and WY.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑