Alaris System PC Unit Model 8015 with software versions 9.17 and 9.19; Central programming, monitoring and power supply component for the Alaris System.
FDA Recall Z-2064-2016 · Posted None
Advertisement
Recalling FirmCareFusion 303, Inc.
Recall StatusTerminated
Product CodeFRN
Date Initiated2016-05-12
Date Terminated2018-12-24
Product Quantity69,693 units
Reason for Recall
A patient weight can be populated incorrectly under certain conditions when using the RESTORE feature to restore infusions running on the Alaris LVP module model 8100 and the Alaris Syringe module model 8110.
Firm Action
An Urgent Medical Device Recall Letter was sent to customers on 5/31/16 to inform them that BD had identified an issue with the Alaris System PC unit model 8015 with software versions 9.17 and 9.19. Firm has received reports that a patient weight can be populated incorrectly under certain conditions when using the RESTORE feature to restore infusions running on the Alaris LVP module model 8100 and the Alaris Syringe module model 8110. The letter informs the customers of the potential risk and the required action for users. Customers are informed that BD will contact them within 60 days to initiate the scheduling process for the software upgrade installation.
Root Cause
Software design
Distribution Pattern
Distributed US (nationwide) and in Canada.