Healon Duet, Part No. 10290080, 10220010, 10220011 and 10220012

FDA Recall Z-2064-2017 · Posted None

Recall StatusTerminated
Product CodeLZP
Date Initiated2017-04-01
Date Terminated2017-11-27
Product Quantity293,867 units total

Reason for Recall

Remote possibility that certain solutions in these lots may contain microscopic glass particles due to damage that occurred at the cylinder neck during the manufacturing process.

Firm Action

A recall letter was sent to customers on 4/1/17 to inform them that AMO has voluntarily initiated this Action because a remote possibility exists that certain Healon OVD solutions in the affected lots may contain microscopic glass particles due to damage that occurred at the cylinder neck during the manufacturing process. Use of OVD solutions with glass particles could potentially lead to intraocular injury. Customers are instructed to complete the Customer Reply Form and fax it to AMO Quality Assurance at 714-247-4510 or email to [email protected] within 3 business days of receipt of the letter. Customers with product complaints or adverse events are instructed to inform AMO by calling 877-266-4543. Customers that do report a complaint are instructed to provide the Healon OVD lot number and, if a patient was involved, the date of surgery, a description of the event and patient outcome.

Root Cause

Process design

Distribution Pattern

US and worldwide: Austria Belgium Croatia Czech Republic Denmark Finland France Germany Great Britain Iceland Ireland Israel Italy Latvia Lebanon Lithuania Netherlands Norway Portugal Spain Sweden Switzerland Tunisia Turkey Russian Fed Guadeloupe Sri Lanka Australia China Hong Kong Indonesia Malaysia Singapore South Korea Taiwan Thailand Japan Chile Colombia Costa Rica Ecuador