Terminated

Therakos UVAR XTS Procedural Kits Is indicated for use in the ultraviolet-A (UVA) irradiation.

FDA Recall Z-2067-2011 · Posted 2011-04-28

Recalling FirmTherakos Inc
Recall StatusTerminated
Product CodeLNR
Date Initiated2011-02-01
Date Terminated2014-03-13
Product Quantity4,323 kits

Reason for Recall

Therakos received an increased rate of complaints of centrifuge bowl leaks for certain lots of Therakos UVAR XTS Procedural Kits (Product Code XT 125).

Firm Action

Therakos notified all direct Therakos US and ex-US End Users and Distributor consignees via Federal Express courier with an Urgent Product Correction Notification, dated 2/1/2011. The letter identified the reason for recall along with the affected product. Customers were to immediately discontinue using and discard the identified affected lots of the product. Customers were to use alternative lots. In order to receive credit for the discarded product, customers were to complete the attached confirmation of receipt and destruction. If there were any questions, a question and answer section was provided but additional questions could be directed towards Therakos Customer Service Center at 1-800-828-6316, Option 2,1.

Root Cause

Process control

Distribution Pattern

Worldwide Distribution

Official Source

View FDA Recall →