LinkSymphoKnee Distal Femoral Augment - Tilastan - Size. 7-8, Lateral-Right/Medial-Left, 5mm-Intended for primary and revision total knee replacement in skeletally mature patients. Product Code: 880-307/21

FDA Recall Z-2067-2023 · Posted 2023-07-03

Recall StatusOpen, Classified
Product CodeJWH
Date Initiated2022-12-23
Product Quantity20 units

Reason for Recall

Fixation screw was inserted through the wrong side of the augment (distal to proximal, as opposed to proximal to distal, which is the correct orientation).

Firm Action

LinkBio Corporation issued A Product Advisory Notice to consignees holding field inventory on 18-May-2023. The product is provided to the surgeon directly by the representative of the Distributor. Letter states reason for recall, health risk and action to take: Please ensure the surgical team is made aware of this issue before using one of the affected augments. While there is no safety impact due to this issue, there could be some confusion during assembly if the issue is not known in advance. No additional action is required by users beyond reinserting the screw in the correct orientation, and no product return is required at this time. Please contact your LinkBio Sales or Customer Service representative or the contact for any questions: Jennifer Lubrecht Senior Director, Quality Management LinkBio Corporation 69 King Street Dover, NJ 07801 Mobile: +1 862.477.0370 Email: [email protected]

Root Cause

Component change control

Distribution Pattern

US Nationwide distribution in the states of AL, IL, IN, LA, GA, NV, TX.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.