OmniTom/OmniTom Elite- X-ray computed tomography applications for anatomy that can be imaged in the 40 cm aperture, primarily head and neck Model Number: NL5000

FDA Recall Z-2068-2023 · Posted 2023-07-03

Recall StatusOpen, Classified
Product CodeJAK
Date Initiated2023-06-06
Product Quantity48 units

Reason for Recall

1. Wheels loosening over time with use of the device could lead to the inability to effectively drive the device and could lead to delay in patient treatment. 2.Batteries used to power the NL5000 device contain contaminated boards, which may impact the ability to power on the device, this could lead to delay for patient treatment

Firm Action

NeuroLogica Corporation issued Urgent Medical Device Correction letter to customers via email June 6, 2023. Letter states reason for recall, health risk and action to take: Users may continue to use the NL5000 system with caution, ensuring they are aware of the potential drive issues due to the wheels. This notice needs to be passed on all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Please transfer this notice to other organizations on which this action has an impact. Please maintain awareness on this notice and resulting action for an appropriate period to ensure effectiveness of the corrective action. NeuroLogica will undertake the voluntary corrective action considering the possibility of delay in patient treatment if the scanner is not operational due to battery failure, as well as the potential for the scanner to halt mid scan. The hardware upgrades provide additional safeguards by improving the design to minimize any risks of the above situations from happening. Service engineers will correct all affected devices free of charge and contact you to arrange for the correction starting June 2023. All corrections in the field will be completed by December 31, 2023. Contact customer service team at 1-888-564-8561 (US/Canada) or 978-564-8561 (International) and [email protected].

Root Cause

Process design

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Australia, Canada, Grand Bahama, India, Italy, Korea, Portugal, Saudi, Scotland, Thailand.