Terminated

ELLIPSE VR, REF: CD1377-36QC Implantable cardioverter defibrillators (ICDs)

FDA Recall Z-2071-2019 · Posted 2019-08-02

Recalling FirmSt Jude Medical Inc.
Recall StatusTerminated
Product CodeLWS
Date Initiated2019-06-20
Date Terminated2024-04-03
Product Quantity47

Reason for Recall

Electrical failures were identified in cardioverter defibrillators (ICDs) due to damaged aluminum wires. ICDs may contain electrical wire connections which may not be completely insulated. The potential patient impact could be the inability to deliver high voltage therapy. There is no available option to verify the vulnerability status for implanted devices.

Firm Action

On 06/20/19, Sales Representatives visited impacted customer accounts to provide talking points that included the reason for recall and to retrieve affected devices. On 06/21/19, hand-delivery of Urgent Medical Device Recall Notices, to physicians supporting implanted patients, commenced. Customers were informed that there is no available option to verify the vulnerability status for implanted devices. Device explant and replacement are recommended. Customers were advised to: 1) Review the device model and serial numbers in the appendix of this letter to identify the impacted patients and return the acknowledgement form to your sales representative; and 2) Device explant and replacement are recommended. The recalling firm will work with customers to provide a replacement device. A copy of this letter is available on https://www.cardiovascular.abbott/us/en/hcp/resources/product/advisories.html. Customers with additional questions were encouraged to call 1-800-727-7846 (Opt3), 8:30am - 5:30pm Central Time, Monday thru Friday.

Root Cause

Process control

Distribution Pattern

Worldwide distribution. US nationwide, and countries: DE, FR, IT, ES, NL, PT, SE, HU, DK, PL, IL, GP, DZ, CH, SA, SK, LU, NZ, VN, and NO

Official Source

View FDA Recall →