TomoTherapy Hi-Art System, ver. 4.0.0, 4.0.1, 4.0.2, ThomoTherapy Incorporated 1240 Deming Way, Madison, WI 53717. The TomoTherapy HI-ART System is intended to be used as an integrated system for the planning and delivery of intensity modulated radiation therapy (IMRT). The HI-ART System provides precise delivery of radiation to tumors or other targeted tissues while minimizing the delivery of radiation to vital healthy tissue.
FDA Recall Z-2072-2010 · Posted 2010-07-22
Recalling FirmTomoTherapy Incorporated
Recall StatusTerminated
Product CodeMUJ
Date Initiated2010-06-23
Date Terminated2011-12-17
Product Quantity29
Reason for Recall
An issue has been identified, as a result of input received from one TomoTherapy Hi-Art Treatment System customer.
With a particular workflow, the Planning Station's Draft and/or final Plan Report can provide incorrect dose statistics.
Firm Action
Consignees were sent on 6/23/10 a TomoTherapy "Urgent Field Safety Notice Medical device Correction" letter dated June 22, 1010. The letter addressed the Issue, Product Affected, Recommended Action and Resolution.
Prior to generating a draft or final Plan Report for a given plan, the plan must be closed and reopened at least once. This issue will be remedied by Hi-Art software version 4.0.3, scheduled for release in late 2010.
Customers with questions were provided with a contact in the letter at the TomoTherapy Customer Interaction Center. The firm can be contacted at 608-824-2800.
Root Cause
Software design
Distribution Pattern
CA, CO, IL, LA, MI, MO, NY, ND, OH, OK, PA,VA, WI, BELGIUM CANADA, KOREA, MALAYSIA, SWITZERLAND, ITALY, GREAT BRITAIN, FRANCE.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source