EOS X-ray Medical Diagnostic , X-ray Equipment System Digital Radiography used in general radiographic examination.
FDA Recall Z-2072-2012 · Posted 2012-09-28
Advertisement
Recalling FirmEos Imaging Inc
Recall StatusTerminated
Product CodeMQB
Date Initiated2012-06-19
Date Terminated2013-09-16
Product Quantity12 EOS Systems installed in the US
Reason for Recall
It was discovered of several generator failures of the EOS System due to failure of its anode controller board that drives the rotation of the X-ray tube anode.
Firm Action
EOS Imaging notified users by letter on 6/19/12 and that an on site correction would be performed by an EOS Imaging Field Service engineer.
EOS indicated in its notification that although this specific defect has not been observed in EOS systems installed in the U.S, they will replace the anode driver boards with new boards equipped with high current capacitors. Similarly, the generator firmware will be upgraded with the software patch that improves the cooling fans' controls. This will be replaced at no charge to the customers, during regularly scheduled preventative maintenance of the user's EOS system. This should be completed by the end of calendar year 2012. Your proposed CAP appears to adequately address the issues you raised in your notification, and is hereby approved.
Your proposed customer notification letter as revised, appears to adequately provide instructions with respect to the use of the product pending the correction of the defect and it appears to provide a clear evaluation in nontechnical terms of the potential hazards related to the defect. Your proposed notification letter is hereby approved for distribution to all affected parties. Further questions please call 678-564-5400.
Root Cause
Radiation Control for Health and Safety Act
Distribution Pattern
Nationwide distribution
Official Source