Allofix Insertion Kit, 2.0 mm, Single Drill, Product Usage: MTF Surgical Instruments are designed for the specific needs of the orthopedic surgeon and are intended to be used in conjunction with the associated allograft. This kit is for single use only.
FDA Recall Z-2075-2015 · Posted 2015-07-15
Advertisement
Recalling FirmMusculoskeletal Transplant Foundation, Inc.
Recall StatusTerminated
Product CodeHXY
Date Initiated2015-05-27
Date Terminated2017-05-08
Product Quantity117 units
Reason for Recall
Musculoskeletal Transplant Foundation is recalling a certain lot of Allofix Insertion Kits because they are labeled as having a 2.0mm diameter drill however it may contain a 2.4 mm diameter drill.
Firm Action
Musculoskeletal Transplant Foundation notified their affected customers of this recall by sending them a Voluntary Recall Notification Letter dated May 27, 2015 or via telephone. The letter identified the affected product, problem and actions to be taken. For questions contact Customer Service at 1-800-433-6576.
Root Cause
Process control
Distribution Pattern
US Nationwide Distribution in the states of AZ, CA, CO, FL, IL, MI, MN, NC, NJ, NV, PA, TN, TX, UT, WA and WV.