Terminated

Navitrack System - OS Knee Universal, Orthopedic Stereotaxic Instrument

FDA Recall Z-2075-2017 · Posted 2017-05-12

Recall StatusTerminated
Product CodeHAW
Date Initiated2008-06-20
Date Terminated2017-05-15
Product Quantity152

Reason for Recall

ZIMMER CAS voluntarily conducted a retrospective recall of the Navitrack System - OS Knee Universal, due to software anomalies which were corrected with a new revision in version 2.2.1.3.

Firm Action

This is a retrospective report of a correction taken on 20 June 2008. The two software issues were corrected in the new software version 2.2.1.3. There were 164 copies of the affected software, 5 of which were only temporary licenses. There were 152 of the 159 software copies that had a full license that were confirmed to have installed the updated instructions per the correction notice for a total percentage of 96%. All sales representatives were made aware of the software issues by written notice. Both software issues were corrected in a new revision of the software with no remaining effect. There have been no complaints since the correction was implemented. Recall notices were e-mailed to customers dated 06/20/2008.

Root Cause

Software design

Distribution Pattern

Domestic: CA, FL, GA, IL, IN, KS, MD, MN, MS, NC, NJ, NY, OR, PA, RI, SD, TN, TX, WA, WI Foreign: Germany, Spain, Czech Republic, Netherlands, Switzerland, New Zealand, Morocco, Brazil, Australia, Austria, France, Italy, Singapore, United Kingdom