U3-3 Solar Blue, Laptop integrated power supply used with the Solar Blue Urodynamics for assessing lower urinary tract dysfunctions
FDA Recall Z-2076-2019 · Posted None
Advertisement
Recalling FirmLABORIE MEDICAL TECHNOLOGIES, CORP
Recall StatusTerminated
Product CodeFEN
Date Initiated2019-07-17
Date Terminated2023-07-10
Product Quantity2 US
Reason for Recall
Power supply unit (PS-SBL) not correctly labeled, and insulation material used at the low voltage DC end does not meet all applicable flammability rating requirements. may lead to a potential fire hazard
Firm Action
Laborie issued letter via FedX on July 17, 2019 advising of the problem, health risk, and action to take: Quarantine the product and unplug the power supply from the wall socket until a replacement is provided. A new power supply unit(s) will be provided along with instructions on how to replace the power supplie in the field. Contact Laborie/MMS customer care with any questions, concerns or problems: Laborie MMS Customer Care
Address: 400 Avenue D, Suite 10, Williston, VT 05495
Phone Number: 800.522.6743
Email: [email protected]
Root Cause
Nonconforming Material/Component
Distribution Pattern
US distribution in AZ, and PA.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.