ColoSense Test Kit, Part No. 80-001, component of ColoSense test
FDA Recall Z-2077-2025 · Posted 2025-07-03
Advertisement
Recalling FirmGeneoscopy, Inc.
Recall StatusOpen, Classified
Product CodeSBB
Date Initiated2025-05-20
Product Quantity10 units
Reason for Recall
Multiple plate failures documented as a result of the Low Positive Template Control being too high and out of range using the affected lot of ColoSense Test Kit.
Firm Action
On May 20, 2025, the firm initiated a voluntary removal of the affected test kit lot. An internal recall notification was sent to the affected CLIA Laboratory Director. All remaining test kits were removed from the CLIA laboratory. The firm's investigation of the root cause for this issue is ongoing.
Root Cause
Unknown/Undetermined by firm
Distribution Pattern
US Nationwide distribution in the state of Missouri.