Siemens syngo.plaza. Radiological image processing system

FDA Recall Z-2078-2012 · Posted 2012-07-26

Recall StatusTerminated
Product CodeLLZ
Date Initiated2012-02-24
Date Terminated2014-01-14
Product Quantity2

Reason for Recall

Siemens became aware of a potential malfunction when using syngo.plaza with software version VA20C_HF01. For datasets with distance measurements it can happen, that all images are not loaded to Viewer but no error message is displayed to user.

Firm Action

Siemens sent a Customer Safety Advisory Notice letter dated February 24, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to ensure that the prefix defined in the viewer settings for "Content Creator's Initials" is not empty. Then all images will be display correctly. This has to be checked for every user. Further questions please call 610-219-6300.

Root Cause

Software design

Distribution Pattern

(USA) nationwide including the states of CA, FL, GA, MA, MN, MO, NJ, NY, OH, WA, and WI.