Terminated

Titan T 5K Motor Product Usage: Low-speed motor is used with various attachments to perform a variety of dental procedures including caries removal, finishing, polishing, pin setting and prophy.The motor has a speed range of 100 to 5000 rpm capability.

FDA Recall Z-2078-2014 · Posted 2014-07-17

Recall StatusTerminated
Product CodeEFA
Date Initiated2014-06-13
Date Terminated2015-02-24
Product Quantity45

Reason for Recall

The housing of 5K motor was laser marked incorrectly as a 20K motor. The recall was expanded in September 2014 to include additional lot #2251181.

Firm Action

Star Dental sent a Medical Device Recall letter, dated June 13, 2014 to affected customers. This letter identified the affected product, problem, and actions to be taken. For question call 866-383-4636 extension 4350. The recall was expanded in September 2014 to include additional lot #2251181. A letter, dated September 2014 was sent to customers who were impacted by the recall expansion.

Root Cause

Process control

Distribution Pattern

US Distribution in the states of : IN, TN, WA, CA, UT, MN, and NY.