Osteonics Scorpio Total Knee; Posteriorly Stabilized Femoral Component; #13 with posts, low friction Ion treatment. Use only with Scorpio Posteriorly Stabilized Tibial Bearing Components with Scorpio Patellar Components.

FDA Recall Z-2079-2008 · Posted 2008-09-16

Recall StatusTerminated
Product CodeJWH
Date Initiated2005-03-17
Date Terminated2008-09-24
Product Quantity1 unit

Reason for Recall

Scorpio Posterior Stabilized Femoral Component labeled and laser marked as a Left, was in fact a Right component.

Firm Action

The International Stryker Distributor was notified via an advisory notice with return receipt on March 18, 2005

Root Cause

Process control

Distribution Pattern

The one unit was distributed to an International Stryker branch.