Disposable StrykeFlow Tip, Stryker Disposable Suction/I irrigator Tip; Sterile, Rx only, Latex Free. Stryker Suction Irrigation system, used to provide a clean and clear operative site through the evacuation and aspiration of irrigant solution. for endoscopic use.

FDA Recall Z-2080-2014 · Posted 2014-07-19

Recalling FirmStryker Endoscopy
Recall StatusTerminated
Product CodeLJH
Date Initiated2014-06-20
Date Terminated2014-10-15
Product Quantity496 boxes (6 tips per box)

Reason for Recall

Internal review found that six lots of Disposable StrykeFlow Tips were shipped with a potentially deformed blister pouch which could impact sterility of the product.

Firm Action

Stryker sent an Urgent Device Removal letter via certified mail on June 20, 2014 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to return any unused product, inform individuals within their organization who need to aware of the device removal. Customers with questions were instructed to call 1-800-624-4422. For questions regarding this recall call 408-754-2664.

Root Cause

Other

Distribution Pattern

Worldwide Distribution - USA (nationwide) and Internationally tAustralia, Canada, France, Greece, Netherlands, United kingdom and South Korea.

Related FDA 510(k) Records