Terminated

Oxoid Antimicrobial Susceptibility Test Discs F300 (Nitrofurantoin), Cat. #CT0036B, packaged in cartridges of 50 discs, 5 cartridges/pack. The firm name on the label is Oxoid Ltd., Basingstoke, Hampshire, England Product is used in the semi-quantitative agar diffusion test method for in-vitro susceptibility testing.

FDA Recall Z-2081-2013 · Posted 2013-08-26

Recalling FirmRemel Inc
Recall StatusTerminated
Product CodeJTN
Date Initiated2013-03-25
Date Terminated2014-06-03
Product Quantity60 packs

Reason for Recall

Use of affected lots may give false indication of susceptibility to Nitrofurantoin.

Firm Action

ThermoFisher sent an Medical Device Recall letter dated March 18, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Please notify any personnel who needs notified of the potential for a false indication of susceptibility to Nitrofurantoin. Please review your inventory for affected product lots and discard the affected product. Return the attached Medical Device Recall Return Response acknowledging the disposal of the affected product. If you have any questions, please contact our Technical Services Department at 800-255-6730 (US) or 913- 888-0939 (International).

Root Cause

Unknown/Undetermined by firm

Distribution Pattern

US Distribution including the states of AL, AR, CA, FL, GA, IL, KS, LA, MA, MD, NY, OK, PA, RI, TX, VA, and WA.