CENTURION Vision System
FDA Recall Z-2081-2016 · Posted 2016-06-28
Advertisement
Recalling FirmAlcon Research, Ltd.
Recall StatusTerminated
Product CodeHQC
Date Initiated2016-05-24
Date Terminated2017-04-13
Product Quantity1 unit
Reason for Recall
The grounding pin within the Power Entry Switch on certain systems was susceptible to detachment when the AC power cable is removed.
Firm Action
Alcon has already notified the affected customer verbally on May 24, 2016, and via written correspondence on June 1, 2016. The device was corrected on May 26, 2016, and the Power Entry Switch that was removed from the customers CENTURION Vision System did not exhibit a grounding pin susceptible to detachment. For further questions, please call (817) 293-0450.
Root Cause
Nonconforming Material/Component
Distribution Pattern
US Distribution to Illinois
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.