Medtronic 0-arm Imaging mobile x-ray system. Model Numbers - B1-700-00027, B1-700-00028
FDA Recall Z-2082-2016 · Posted None
Advertisement
Recalling FirmMedtronic Navigation, Inc.-Littleton
Recall StatusTerminated
Product CodeOXO
Date Initiated2016-03-17
Date Terminated2021-04-06
Product QuantityUS - 581
Reason for Recall
It has been discovered that replacement x-ray generator and system motion batteries could become damaged during shipment, resulting in unacceptable risk of failure.
Firm Action
Medtronic's planned action to bring device into Compliance.
1. You will provide an Errata sheet to customers updating the missing and incorrect information.
2. Field Service Engineers (FSEs) will perform radiation output measurements via validated procedures to attain compliance with
21CFR part 1020.32(k)(6).
3. You will inspect and test X-ray generator and system motion batteries for damage, in accordance with the plan submitted to CDRH. Damaged batteries will be replaced.
4. You will update the systems software-based instructions for use by a software update which includes the information supplied by the Errata sheet.
For further questions, please call (800) 595-9709.
Root Cause
Radiation Control for Health and Safety Act
Distribution Pattern
Nationwide Distribution