Terminated

IGuide System, software versions: 2.2.0, 2.2.1, 2.2.2 Product Usage: Linear Accelerator. The intended use of the device is the control of accurate patient positioning with assistance of a 30 Tracking System in a radiotherapy environment.

FDA Recall Z-2082-2019 · Posted None

Recall StatusTerminated
Product CodeIYE
Date Initiated2019-07-18
Date Terminated2023-08-11
Product Quantity496 units

Reason for Recall

Sometimes, although the HexaPOD has reached the target position, the Constant Correlation Check (CCC) can be too sensitive and iGUIDE may incorrectly demand a verification scan.

Firm Action

The firm sent an Urgent Important Field Safety Notification letter dated July 2019 via email. The identified the affected product, problem and actions to be taken. Customers were instructed to post this notice in a place accessible to all users, e.g. Instructions for Use, until this action is closed and advise the appropriate personnel, working with this product, on the content of this letter.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution - US Nationwide

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →