Medtronic Navigation O-arm 02 surgical x-ray imaging system
FDA Recall Z-2083-2016 · Posted None
Advertisement
Recalling FirmMedtronic Navigation, Inc.-Littleton
Recall StatusTerminated
Product CodeOWB
Date Initiated2016-03-17
Date Terminated2021-04-06
Product QuantityUS - 63
Reason for Recall
Review of on-system electronic product labeling and Instr uctions for Use in response to an internal management concern uncovered gaps in compliance with the requirements for information to be provided to users in 21CFR part 1 020.30(h).
Firm Action
Medtronic planned action to bring device into Compliance:
1. You will contact customers and provide an Errata sheet to them updating the missing and incorrect information.
2. Medtronic field service personnel will perform radiation output measurements via validated procedures to attain compliance with 21CRF part 1020.32(k)(6).
3. You will update the systems software-based instructions for use by a software update which includes the information supplied by the Errata sheet.
For questions call (800)-595-9709.
Root Cause
Radiation Control for Health and Safety Act
Distribution Pattern
Nationwide Distribution
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.