RX Daytona, Model Nos. RX4040 and RX4041
FDA Recall Z-2083-2019 · Posted None
Advertisement
Recalling FirmRandox Laboratories, Limited
Recall StatusTerminated
Product CodeJJE
Date Initiated2019-06-24
Date Terminated2020-05-21
Product Quantity3
Reason for Recall
There was a report of a malfunction with the sensor which resulted in an increase of temperature leading to melting of internal wires and scorch damage to the internal structure of the instrument.
Firm Action
On June 24, 2019, the firm notified its customers of the recall through an Urgent Field Safety Notice.
Customers were asked to take the following actions:
If there is any sign of damage to the RCU lid or lid sensor
" Switch of main analyser main power supply as soon as possible
" Contact technical support immediately.
" Discuss the contents of this notice with your Medical Director.
" Display this notice in a prominent location near the instrument.
" Complete and return the response form 12187-QA to [email protected]
within five working days.
Root Cause
Under Investigation by firm
Distribution Pattern
Domestic distribution to CA CT MA MD MO NC NJ NY OH PA UT WV and PR. Worldwide foreign distribution.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.