Patient Handling System (Motion Control Software), Product Usage: Indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body when radiation treatment is indicated.

FDA Recall Z-2085-2015 · Posted 2015-07-16

Recalling FirmViewray Incorporated
Recall StatusTerminated
Product CodeIYE
Date Initiated2015-04-01
Date Terminated2016-02-04
Product Quantity3

Reason for Recall

ViewRay received a report that the couch moved unexpectedly into the bore after performing a RTCS reboot.

Firm Action

ViewRay sent a Customer Advisory Notification Letters dated April 1, 2015 to thier customers. The letter identified the affected product, problem and actions to be taken. For questions or concerns regarding this issue, contact ViewRaylM Customer Support at [email protected] or call the ViewRaylM Customer Support center at 855-286-8875.

Root Cause

Software Design Change

Distribution Pattern

US Nationwide Distribution in the states of CA, MO & WI