026-20036A Liberty Drain Line, Luer Tap For patients with acute and chronic end stage renal disease going under PD in healthcare or home setting
FDA Recall Z-2085-2021 · Posted None
Advertisement
Recalling FirmFresenius Medical Care Holdings, Inc.
Recall StatusTerminated
Product CodeFKX
Date Initiated2021-04-28
Date Terminated2022-08-08
Product Quantity6 Cases
Reason for Recall
Exposed to freezing temperatures due to refrigerated truck malfunctions
Firm Action
Fresenius issued telephone script communication on April 28.2021 stating reason for recall, heath risk and action to take:
Tell them you already contacted their Home Therapy Nurse to alert them that you would be calling them.
o The purpose of my call today is that you received a delivery from us on (either Feb 22nd or Feb 23rd).
o The truck you received a delivery from had a temperature malfunction due to the bad storms Texas experienced back in February. This caused the product you were delivered to be exposed to freezing or near freezing temperatures for about 3 hours.
o
o I see you received on .
o (if received any PD SOLUTION/SALINE/or CONCENTRATES):
Explain that the solutions inside their product are not affected and that we have freeze/thaw studies that prove this.
Explain that due to the freezing temperatures, the packaging may become compromised, and they may have seen leaks.
For the other products such as Hand Sanitizer and other ancillary products, again, the product could be compromised according to their packaging requirements listed on their packaging.
o We are giving you the opportunity to return these products if you choose to do so at this time.
o If returning product(s):
o Fill out RGA spreadsheet with customer information including how much product they are returning (include lot# and amount returning).
Driver will pick up returned products on their next scheduled visit.
Fresenius Corrections and Removals Rep: Email Post Market Surveillance to open a complaint against product.
o If not returning product(s):
o Thank them for their time and reiterate to always inspect their packaging for leaks as this is what they are trained to do anyways from their Home Therapies Nurse.
If you have any questions call 781-699-2243 or email: [email protected]
Root Cause
Environmental control
Distribution Pattern
US Distribution to: TX only