TRIGEN Hind Foot Fusion Nail, 10mm X 16mm Left, REF 71701016L. The device is indicated for degeneration, deformity, or trauma of both the tibiotalar and talocalcaneal articulations in the hindfoot.

FDA Recall Z-2086-2008 · Posted 2008-09-16

Recalling FirmSmith & Nephew Inc
Recall StatusTerminated
Product CodeHSB
Date Initiated2008-04-22
Date Terminated2010-05-28
Product Quantity40

Reason for Recall

One of the distal locking screw holes on a hindfoot fusion nail was drilled with an incorrect trajectory.

Firm Action

All affected Smith & Nephew Sales Representatives were notified of problem and the recall via certified letter and e-mail on 4-22-08. All affected Smith & Nephew International Distributors were notified via email on 4-22-08 and 4-23-08. They were instructed to immediately notify their accounts concerning the recall. If you have questions, contact Customer Service at 800-238-7538.

Root Cause

Process control

Distribution Pattern

Nationwide and Australia, Belgium, Canada, Denmark, France, Germany, Italy, The Netherlands, Norway, South Africa, Spain, Sweden, and United Kingdom.