Cadence Size 5 Left Tibial Tray Implant, designed to treat ankle arthritis through replacement of the ankle joint with a prosthesis, thereby reducing pain, restoring alignment, and allowing for movement at the replaced joint. Catalog Number: 10207205; Lot Number #QJ0248
FDA Recall Z-2086-2016 · Posted None
Advertisement
Recalling FirmIntegra LifeSciences Corp.
Recall StatusTerminated
Product CodeHSN
Date Initiated2016-04-29
Date Terminated2016-10-19
Product Quantity12 units
Reason for Recall
As a result of an internal review of all labeling for the Cadence System, it was determined that the outer package label for the Size 5 Left Tibial Tray Implant reflects the material composition incorrectly. The labeling incorrectly reflects the material as Cobalt Chromium instead of Titanium.
Firm Action
On 4/29/16, US consignees were notified by electronic mail delivered by traceable method.
Root Cause
Labeling mix-ups
Distribution Pattern
NC and PA
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.