Cadence Size 5 Left Tibial Tray Implant, designed to treat ankle arthritis through replacement of the ankle joint with a prosthesis, thereby reducing pain, restoring alignment, and allowing for movement at the replaced joint. Catalog Number: 10207205; Lot Number #QJ0248

FDA Recall Z-2086-2016 · Posted None

Recall StatusTerminated
Product CodeHSN
Date Initiated2016-04-29
Date Terminated2016-10-19
Product Quantity12 units

Reason for Recall

As a result of an internal review of all labeling for the Cadence System, it was determined that the outer package label for the Size 5 Left Tibial Tray Implant reflects the material composition incorrectly. The labeling incorrectly reflects the material as Cobalt Chromium instead of Titanium.

Firm Action

On 4/29/16, US consignees were notified by electronic mail delivered by traceable method.

Root Cause

Labeling mix-ups

Distribution Pattern

NC and PA