Hettich Centrifuge Modules: StreamLAB Centrifuge Module (10372372 or 10444807 or 10706836) Aptio Centrifuge Module (10703032 or 10713761) FlexLab High Throughput Centrifuge Module (11171747) FlexLab and Aptio Centrifuge Module (10628157 or 11266581): Used to separate substances according to their densities.

FDA Recall Z-2087-2016 · Posted None

Recall StatusTerminated
Product CodeLXG
Date Initiated2016-05-11
Date Terminated2018-04-27
Product Quantity487

Reason for Recall

Premature bucket failures

Firm Action

Siemens mailed to customers an Urgent Medical Device Correction (UMDC#LAI16-02.A.US) Urgent Field Safety Notification letter informing customers of the problem with recommendations to follow the existing maintenance instructions in the product operator manuals as well as a new requirement to spin at cycle speeds not exceeding 3500 rpm.

Root Cause

Under Investigation by firm

Distribution Pattern

Nationwide: AL, AR, AZ, CA, CO, CT, FL, GA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, PA, SC, SD, TN, TX, VA, WA, WI, WV.