Hemochron Jr. Whole Blood Microcoagulation System APTT Cuvette. For in vitro diagnostic coagulation test. Catalog number J103.

FDA Recall Z-2088-2014 · Posted 2014-07-21

Recalling FirmITC-Nexus Dx
Recall StatusTerminated
Product CodeJPA
Date Initiated2014-06-20
Date Terminated2014-12-11
Product QuantityUS- 315 cuvettes; OUS - 2835 cuvettes

Reason for Recall

Test results may be erroneously elevated or report an out-of-range high (OOR-H) error code in non heparinized blood samples.

Firm Action

ITC issued an Urgent Medical Device Recall Letters/Response Forms dated June 20, 2014, to End-users and US Distributors were notified. ITC asks that this communication be forwarded to all those within your organization who need to be aware of this matter. If any affected products have been forwarded to another facility, a copy of this communication should be forwarded to them immediately. Customers should check their inventory for the affected lot. The affected lot should not be used and quarantined until return to ITC. The attached form should be completed and returned to ITC within 10 days. ITC will review the form and contact the customer to coordinate any returns. For questions, please contact Technical Support: Telephone 732-548-5700, Extension 4011 E-mail: [email protected] Normal business hours are Monday through Friday, 8:00 AM to 6:30 PM, Eastern Time.For urgent matters, Technical Support is available 24 hours per day, 7 days a week.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution - US including the states of Georgia, Virginia and New York., and the countries of Canada, France, Italy, Liechtenstein and Romania.