TrueDent Clear, TDM 100, Model No. OBJ-09164 Light-curable methacrylate-based resin that enables fabrication of dental appliances

FDA Recall Z-2090-2025 · Posted 2025-07-07

Recalling FirmSTRATASYS LTD
Recall StatusCompleted
Product CodeEBI
Date Initiated2024-01-19
Product Quantity78

Reason for Recall

Customers unable to use cartridges due to formatting error in expiration date which leads to switching the day with the month. Or, the system will not recognize that a resin is expired and will not prevent the user from using an expired material.

Firm Action

The firm initiated the recall to the consignee level on January 15, 2024. Customers were asked to coordinate a cartridge replacement. The reporting of this recall was done by the firm retroactively; the issue was resolved in January 2024.

Root Cause

Software design

Distribution Pattern

US Nationwide distribution including in the states of Arizona, Arkansas, California, Colorado, Connecticut, Florida, Georgia, Illinois, Indiana, Maryland, Massachusetts, Michigan, Nebraska, New York, North Carolina, North Dakota, Oklahoma, Oregon, Utah, Wisconsin.