Disposable Replacement Backflush Brush Needles (23 gauge/0.6mm)-for use in vitreoretinal surgery Product Number: labeled as 1281.B1D, product is 1281.B1D06
FDA Recall Z-2091-2021 · Posted None
Advertisement
Recalling FirmDutch Ophthalmic USA, Inc.
Recall StatusTerminated
Product CodeHMX
Date Initiated2021-02-25
Date Terminated2021-11-05
Product Quantity12 US
Reason for Recall
Mislabeled: 1281.B1D Disposable Replacement Backflush Brush Needles label states (20 gauge/0.9 mm) instead of the correct size of 23 gauge/0.6 mm and the correct product code 1281.B1D06
Firm Action
Dutch Opthalmic issued Urgent Medical Device Recall letter on 2/25/21 stating reason for recall, health risk and action to take:
We request that you take the following immediate actions:
Verify whether you have unused boxes or individual blisters of LOT 3420011396 in your inventory and quarantine product subject to this reca ll.
Contact Customer Service at our distributor Dutch Ophthalmic, USA at 1-800-753-8824 x 1 to receive a Return Material Authorization (RMA) number and receive instructions on how to return the unused affected product.
Complete the attached Device Recall Reconciliation Form, sign it and return a scan or photo of the signed form by email to [email protected].
Add the original completed and signed Device Recall Reconciliation Form
referencing the Return Material Authorization number with the product to be
returned. In the next few days a Service rep will contact you to discuss replacement of returned products. Contact Dutch Ophthalmic 1-800-753-8824
x1 for assistance.
Root Cause
Under Investigation by firm
Distribution Pattern
Distribution in US states of IL, IN, MO, PA.
Official Source