VERSA-FX FEMORAL FIXATION SYSTEM COMPRESSION LAG SCREW, several length (55 mm to 155 mm), diameters (12.7 mm and 15.8 mm) and short thread, standard thread and large thread. Item codes starting with 00-1193-005, 00-1193-010 and 00-1193-015. orthopedic implant.

FDA Recall Z-2092-2013 · Posted 2013-08-26

Recalling FirmZimmer, Inc.
Recall StatusTerminated
Product CodeKTT
Date Initiated2013-06-19
Date Terminated2015-06-16
Product Quantity85, 794 units in total

Reason for Recall

Single packaging configuration used for the VERSA-FX-II Femoral Fixation System failed the leak test compromising the sterility of the product.

Firm Action

The firm, Zimmer, sent an "URGENT MEDICAL DEVICE NOTIFICATION" letter dated July 11, 2013 to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to inspect the device according to the insert instructions; not use the device if damage to the corner of the inner tray is observed, and contact the firm for return and questions. Customers should also keep this notification in mind and inform Zimmer of product with this condition using [email protected]. If after reviewing this notification you have questions or concerns, please contact Zimmer at 1-877-946-2761.

Root Cause

Packaging process control

Distribution Pattern

Worldwide distribution: US Nationwide (including PR and USVI), and countries of: Israel, Argentina, Brazil, France, Mexico, Saudi Arabia, Beirut, China, Korea, San Salvador, Honduras, Canada, Australia, Switzerland, Germany, Arab Emirates, UK, Belgium, Egypt, Spain, France, Italy, Iran, Lebanon, Jordan, Iraq, Canary Islands.