Terminated

Boston Scientific LeVeen SuperSlim Needle Electrode, 3.0 cm/15 cm, sterile. The device is intended to be used in conjunction with a Radio Therapeutics Corporation radiofrequency (RF) generator for thermal coagulation necrosis of soft tissue, including partial or complete ablation of nonresectable liver lesions.

FDA Recall Z-2093-2008 · Posted 2008-09-16

Recall StatusTerminated
Product CodeGEI
Date Initiated2007-10-01
Date Terminated2008-09-16
Product Quantity15,604 of all products

Reason for Recall

The cannula may become detached from its correct orientation inside the handle and may prevent retraction of the tines.

Firm Action

Customer notifications were sent to customers in Japan on October 1, 2007 instructing them to check their inventory and to return affected product. Contact Rob Miragliuolo at 508-683-4186 if you have questions.

Root Cause

Device Design

Distribution Pattern

Japan.