Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

VERSA-FX II FEMORAL FIXATION SYSTEM COMPRESSION TUBE/PLATE, standard and short - low profile, different hole/length combinations (4/86, 5/102 and 6/118), different degree angles (130, 135, 140, 145 and 150). Item codes starting with 00-194-130, 00-1194-135, 00-1194-140, 00-1194-145, 00-1194-150)., orthopedic implant.

FDA Recall Z-2094-2013 · Posted 2013-08-26

Recalling FirmZimmer, Inc.
Recall StatusTerminated
Product CodeKTT
Date Initiated2013-06-19
Date Terminated2015-06-16
Product Quantity85, 794 units in total

Reason for Recall

Single packaging configuration used for the VERSA-FX-II Femoral Fixation System failed the leak test compromising the sterility of the product.

Firm Action

The firm, Zimmer, sent an "URGENT MEDICAL DEVICE NOTIFICATION" letter dated July 11, 2013 to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to inspect the device according to the insert instructions; not use the device if damage to the corner of the inner tray is observed, and contact the firm for return and questions. Customers should also keep this notification in mind and inform Zimmer of product with this condition using [email protected]. If after reviewing this notification you have questions or concerns, please contact Zimmer at 1-877-946-2761.

Root Cause

Packaging process control

Distribution Pattern

Worldwide distribution: US Nationwide (including PR and USVI), and countries of: Israel, Argentina, Brazil, France, Mexico, Saudi Arabia, Beirut, China, Korea, San Salvador, Honduras, Canada, Australia, Switzerland, Germany, Arab Emirates, UK, Belgium, Egypt, Spain, France, Italy, Iran, Lebanon, Jordan, Iraq, Canary Islands.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑