Terminated

AMO 25GA Vitrectomy Cutter (Model Number NGP0025) and AMO 20GA Vitrectomy Cutter and Irrigation Sleeve (Model Number NGP0020) Used with the AMO Whitestar Signature System. For ophthalmic use.

FDA Recall Z-2096-2017 · Posted 2017-05-16

Recall StatusTerminated
Product CodeMLZ
Date Initiated2017-03-30
Date Terminated2017-08-23
Product Quantity360 units total

Reason for Recall

Packaging error. 20GA Vitrectomy Cutters may be found in a 25GA package and 25GA Vitrectomy Cutters may be found in a 20GA package.

Firm Action

An Urgent Field Safety notice was sent on 3/30/17 to inform customers that AMO has voluntarily initiated this Action because a possibility exists that, due to a packaging error, 20GA Vitrectomy Cutters may be found in a 25GA package and 25GA Vitrectomy Cutters may be found in a 20GA package. Use of a Vitrectomy Cutter that is a different size than expected could lead to the need to alter the surgical technique, including conjunctival dissection, incision enlargement and scleral sutures.

Root Cause

Packaging process control

Distribution Pattern

Worldwide distribution to SA, TR, AZ, NL, PT, SG, AU, JP, and HK.