Proton therapy system -Proteus 235 aka Proteus Plus. A medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.
FDA Recall Z-2097-2017 · Posted 2017-05-03
Advertisement
Recalling FirmIon Beam Applications S.A.
Recall StatusTerminated
Product CodeLHN
Date Initiated2017-04-10
Date Terminated2018-02-16
Product Quantity2
Reason for Recall
A PBS (Pencil Beam Scanning) irradiation may pause for different reasons. After a pause, irradiation may restart from the beginning instead of recalculating the field based on the already delivered dose.
Firm Action
Target date for software update June 30, 2017.
A Field Safety Notice was issued on 4/10/2017 describing the problem and reminding users to follow guidance in the user's manual.
Root Cause
Software design
Distribution Pattern
FL