KLS Martin Battery Pack, REF: KLS BP 001, and KLS BP 040 (bulk case of 40), used in conjunction with the KLS Martin MaxDriver
FDA Recall Z-2098-2023 · Posted 2023-07-05
Advertisement
Recalling FirmPro-Dex Inc
Recall StatusOpen, Classified
Product CodeGEY
Date Initiated2023-02-15
Product Quantity2000
Reason for Recall
Battery packs used in surgical procedures where a surgeon uses a powered screwdriver and a right-angle attachment have switched manufacturing and expiration dates on sterile pouch labels, so batteries will appear to be prematurely expired.
Firm Action
On 2/15/23, a recall notice was sent to a customer asking them to cease distribution and return affected product to the recalling firm.
Questions can be directed to 949-769-3200.
Root Cause
Error in labeling
Distribution Pattern
US: FL
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.