NEFA CAL (Non-Esterified Fatty Acids)- IVD Standard for measurement of NEFA Catalog Number: FA115

FDA Recall Z-2099-2021 · Posted None

Recall StatusTerminated
Product CodeJLH
Date Initiated2021-05-10
Date Terminated2021-11-10
Product Quantity71 kits US

Reason for Recall

Standard packed within NEFA kit has been assigned incorrectly. Calibrating with the incorrect value will cause a positive shift of up to +8% with both quality controls and patient samples

Firm Action

Randox notified the distribution center within the USA via email on 5/10/21.The distributor will then contact the customer directly. Letter states reason for recall, health risk, and action to take: Action to be taken: " Please discard previous value sheets and ensure standard value is updated. " Review results generated with the affected batches in line with the clinical profile of the patient. " Discuss the contents of this notice with your Medical Director

Root Cause

Under Investigation by firm

Distribution Pattern

US Nationwide Distribution in the states of AR, CA, CT, NY, OH, PA, TN, and TX